Description
Supply and recovery switch over system for cryogenic tanks with medical oxygen.
This device operates so that gas consumption is optimized with the least possible waste of product, alternately switching the two storage sections and recovering the excess of gas phase redirecting it to the outlet line. All this is done automatically and with the possibility of a remote control in telemetry.
- Real time monitoring measures and pressures through LCD display with backlight
- Alarm signaling through blinking visual and acoustic indicators with recovery time programmable by the user
- Programming by encoder with button integrated
- Alarm acoustic signal with acknowledge pushbutton on the front panel
- Solenoid valves status indication on front panel
- Automatic switch over functions for empty tank, extra consumption or fault storage, through integrated solenoid valves
- Recovery function for gaseous fraction produced by the evaporation of the backup tank, through integrated solenoid valves
- Alarm or shut-off function for low temperature detection on operating tank supply
- Set points programmable by the user for minimum level, high and low operating tank pressures, high and low supply output pressure, low III° source pressure
- Proprietary network protocol CRIONet
- Remote programming and reception of update messages through CrioSystemSupervisor platform (MOD-COM modem required)
- Auxiliary output switch for external cumulative alarm indicator or hooter
- Visual and acoustic indication about alarm status in accordance with European standard ISO 7396-1 and EN 60601-1-8