Description
SDAM customizable devices allows to control gas distribution inside the Hospital facility from the storage to the patient.
Monitoring of 2nd stage pressure regulators for medical gases
- 5 digital alarm inputs ON/ OFF (pressure switches) with the following configuration:
- Low/ High alarm pressure: O2, Med. Air and Low Vacuum or Evac. Air, CO2 and
- Low Vacuum
- Alarm notification by 5 red LEDs
- Acknowledge button for acoustic sound with alarm recovery function setting by a dip-switch (max. 15 min)
- Test button for acoustic and visual periodic check
- Programming jumper for inputs operation N.O. or N.C.
- Local network interface with protocol SDAMNet Plus or ModBus (option)
- Network address programming by a dip-switch
- Auxiliary output switch for external cumulative alarm indicators
- Auxiliary output switch for external hooters
- 5 independent outputs reproducing the inputs status
- Visual and acoustic indications about alarm status in accordance with European standard ISO 7396-1 and EN 60601-1-8
Data sheet download










Ambra Sistemi S.r.l.