SDAM-MED D10

 

Features

Monitoring of 2nd stage pressure regulator units in medical gases delivery systems
Monitoring of 1st stage medical gas storages

  • 10 digital inputs for on/off contacts for pressure switches and vacuum switches
  • Alarms and status monitoring with associated ID
  • Classification as Emergency or Operative alarm for each input
  • Display with 2 lines x 20 characters with LED backlight
  • Yellow and red alarm lights associated to Emergency or Operative alarm status
  • Acknowledge button for acoustic sound with alarm recovery function
  • Auxiliary output switches for external cumulative alarm indicators or hooters
  • Proprietary network protocol SDAMNet Plus or ModBus
  • Programming by Windows software application SDAM Programming Tool
  • Optical and acoustic indications about alarm status in accordance with European standard ISO 7396-1 and EN 60601-1-8

Technical Specs

REGISTER

Register to view technical data and to download data sheets.

 

Power supply
230 Vac 50/60 Hz., Pmax. 5 VA, fuse 0.5 A
Visual indications
Display 2 lines x 20 characters with backlight
Operative alarm: LED Ø 5 mm, yellow
Emergency alarm: LED Ø 5 mm, high efficiency, red
Alarm sound
Piezoelectric speaker, volume adjust up to 85 dB @ 1 m
Inputs
10 self-powered digital inputs for switch sensors, photo-coupled, maximum current 15 mA for each input, supply voltage 20 Vdc

Auxiliary output switch
V max = 48 Vdc/Vac, Imax = 1 A, N.C. or N.O. configuration by software

Power output 
Low voltage for optional RF Module (LoRa radio)

Housing
Plastic 9 modules DIN rail (EN 60715) in accordance with DIN 43380, PVC, dim. 160x90x(h)58 mm, IP34
Local network interface
Isolated bidirectional RS485 protocol SDAMNet Plus or ModBus
Weight
560 g
Operating temperature
-10…50 °C
Reference standards and directives
93/42/EEC, 2014/30/EU EMC, 2014/35/EU LVD, EN60601-1,
ISO 7396-1, EN 60601-1-8, EN61000-6-1, EN61000-6-3, RoHs 2011/65/EU

 


Typical Applications

Alarm unit for on/off contacts


CE 0425
Compliant with 93/42/EEC
Medical Devices Directive - Class IIb

TECHNICAL FEATURES

Power supply 230VacAcoustic alarm Visual alarm  