Monitoring of 2nd stage area gas shut-off valves for medical gases

  • 4 digital inputs for position valve sensors, type NAMUR or ON/ OFF
  • Valve status by light indicators (4 green LEDs = valve open and/ or 4 red LEDs = valve close)
  • Valve fault indication (green LEDs/ red LEDs alternating) only for NAMUR version Test button
  • Inputs programming operation by dip switches
  • Configuration function as base unit or repeater unit
  • RS485 local SDAMNet Plus network or Modbus protocol (optional)
  • Network address programming by dip-switch
  • Auxiliary outputs for collective valves open
  • Auxiliary outputs for collective valves closed
  • 4 independent outputs reproducing the input status (N.C. logic)
  • Gas labels available: O2, N2O, Med. Air 8 bar, Med. Air 4 bar, Evac. Air, Compr. Air, Air 800, CO2, Vacuum, N2, He

 

Technical Specs

REGISTER

Register to view technical data and to download data sheets.

HOSPITAL ALARM


Typical Applications

Local repeater for networks constituted by SDAM-MED and/or SDAL-MED 

CE 0425
Compliant with 93/42/EEC
Medical Devices Directive - Class IIb

TECHNICAL FEATURES

Power supply 230VacAcoustic alarmVisual alarm   

 


 