AMBRA SISTEMI



We'd like to hear your message!

Contact us

SEND MESSAGE

CRIOTEL LCD

 

Features

  • Local and remote monitoring of cryogenic tank (vertical and horizontal) for level and pressure monitoring
  •  Integrated and patented differential and relative pressure transducers
  •  Large LCD display for level and pressure visualization with backlight activated by low light sensor
  •  Level to % conversion pushbutton (if another unit of measurement is configured) present on the front panel
  •  Minimum level, low pressure and high pressure/high level set-points with additional output switches programmable by user
  •  Low temperature alarm for detecting leakages of cryogenic media
  •  Additional alarm inputs for external switch sensors (e.g. pressure switches)
  •  Repeater signals 4…20 mA or 0…20 mA outputs for level and pressure measures
  •  Output signals for low level, low pressure, high pressure or high level
  •  Proprietary network protocol CRIONet
  •  Programming by rotary encoder with button integrated
  •  Remote programming and reception of update messages through CrioSystemSupervisor platform (MOD-COM modem required)
  •  Easy, fast and inexpensive installation for setup and startup
 

Technical Specs

REGISTER

Register to view technical data and to download data sheets.


Power supply
24 Vdc ±5 %, max 0.2 A / 230 Vac, max 100mA (optional)
Pressure and level (DRT transducer)
Relative pressure:
Range: 0...20 bar (standard); 0...50 bar (optional)
Resolution: 0.1 bar
Combined error (0...50 °C): < ±0,2 bar
Breaking overpressure: 35 bar (80 bar for range 0...50 bar)

Differential pressure (level):
Range: 0-2 bar
Resolution: --- (programmable by user)
Combined error (0...50°C): < ±2 % of F.S @ F.S. ≥150 mbar
Differential overpressure: ± 4 bar limited by integrated protection system
Additional alarm inputs
3 digital inputs, 9-48 Vdc, Imax 5 mA, insulation >4 kV
Output signals
2 independent 4...20 mA sourcing outputs for level and pressure, open circuit voltage >16V, maximum load 500 Ω, adjustable range to 0...20 mA or 4...20 mA

3 Polarized output signals, solid state, Vmax 48 Vcc, Imax 100 mA, insulation >4 kV

Visual indications
4 digits LCD display for level H18 mm
4 digits LCD display for pressure H18 mm
3 red led Ø 5 mm for additional alarms
1 yellow led Ø 5 mm for data communication, 1 red led Ø 5 mm for unit fail
Pneumatic connections
1/8” G female with fitting adapters 1/4" NPT female or 1/4" G female
Housing
White aluminum enclosure for wall mounting, size 200 x 170 x 90 mm, waterproof degree IP 65
Local network interface
Proprietary protocol CRIONet on insulated RS485
Weight
3,5 kg
Operating Temperature
-20...60 °C
Reference standards and directives
2014/30/EU EMC, 2014/35/EU LVD
EN 61000-6-1, EN 61000-6-3, ISO 15001, BAM M 034-1


Typical Applications


Local and remote monitoring unit for cryogenic tanks

TECHNICAL FEATURES

Power supply 230VacPower supply 24VccVisual alarm
Outputs 4-20Level sensorPressure_sensorTemperature sensor
CRIONet 485 interfaceMouting with magnetsCE certification                        

Haven't found the right solution for your needs?

Do you have any questions or do you need clarification? We are happy to answer your questions at any time.

contact us

Documents download


Download PDF Data sheet