HOSPITAL ALARM

SDAL-MED L9

SISTEMA ALLARMI DIGITALI
 

Download PDF Data sheet

Download PDF Conformity certificate
 

Typical application:
-Monitoring of 2nd stage pressure regulators for medical gases


Alarm unit SDAL-MED L9 functions:
•    9 digital alarm inputs ON/ OFF (pressure switches) with the following configuration:
     -L/ H alarm pressure O2, N2O, Med. Air, Air 800/ N2 and Low Vacuum
•    Alarm notification by 9 red LEDs
•    Acknowledge button for acoustic sound with alarm recovery function setting by a dip-switch (max. 15 min)
•    Test button for acoustic and visual periodic check
•    Programming jumper for inputs operation N.O. or N.C.
•    Local network interface with protocol SDAMNet Plus or ModBus (option)
•    Network address programming by a dip-switch;
•    Auxiliary output switch for external cumulative alarm indicators
•    Auxiliary output switch for external hooters
•    9 independent outputs reproducing the inputs status
•    Visual and acoustic indications about alarm status in accordance with European standard ISO 7396-1 and EN 60601-1-8.

Functions:
- No.9 digital inputs ON/ OFF (pressure switch) with the following configuration:
  L/ H alarm pressure O2, N2O, Med. Air,   Air 800/ N2 and Low Vacuum 
- Alarm notification by 9 red Led for high priority
- Alarm recovery function after 15 minutes, selectable by dip-switch
- Test button
- Programming jumpers for selection of input operation N.O. or N.C.
- Auxiliary output switch for external cumulative alarm indicator
- Auxiliary output switch for external hooter
- No.9 independent outputs reproducing the input status
- RS485 local network SDAMNET or Modbus protocol (option)
- Network address programming by dip-switch
- Optical and acoustic indication about alarm status in accordance with European standard ISO 7396-1 and IEC 60601-1-8